Recall Z-0478-2026 ยท NKB Thoracolumbosacral Pedicle Screw System

Orthofix U.S. LLC

Virata Preassembled Polyaxial Screw, utilized in the Virata Spinal Fixation System. The following sizes are affected: 1. POLYAXIAL SCREW, CC, 5.5mm X 30mm; Item Number: TL1-10755030. 2. POLYAXIAL SCREW, CC, 5.5mm X 35mm; Item Number: TL1-10755035. 3. POLYAXIAL SCREW, CC, 5.5mm X 40mm; Item Number: TL1-10755040. 4. POLYAXIAL SCREW, CC, 5.5mm X 45mm; Item Number: TL1-10755045. 5. POLYAXIAL SCR

DateNov 7, 2025 (initiated Sep 19, 2025)
StatusOpen, Classified
ReasonPolyaxial screws do not meet performance standards.
Root causeProcess change control
ActionOrthofix notified consignees via email on 09/19/2025. Consignees were instructed to immediately discontinue use of the affected devices, quarantine all affected inventory, and return units to Orthofix using the provided Return Goods Authorization (RGA) process. Surgeons were instructed to monitor affected patients according to standard clinical practice. Customers were advised not to implant or distribute any remaining affected product. All consignees were requested to complete and return the response form.
Quantity624 units
DistributionUS Nationwide distribution in the states of California, North Carolina, and Texas.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K250908Virata Spinal Fixation System
SeaSpine Orthopedics Corporation, Jun 18, 2025