Recall Z-0163-2026 ยท LNH System, Nuclear Magnetic Resonance Imaging
IMPRIS Imaging Inc
IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
| Date | Oct 28, 2025 (initiated Sep 30, 2025) |
| Status | Open, Classified |
| Reason | There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. |
| Root cause | Under Investigation by firm |
| Action | Imris issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 9/30/2025 via email. The notice explained the issue, the risk to health, and provided the following information: "If the error message, "Magnet Supervision: The average heater power is above the alarm threshold. Please contact Siemens Service." appears, DO NOT REBOOT THE SYSTEM. Please call your local Siemens Healthineers service organization. Grant access to the system so that the inspection can be carried out as directed by Siemens Healthineers service organization." "Please immediately instruct your person |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China. |
| FDA record | Open on fda.gov |