ClearanceBrief
Recall Z-0161-2026 ยท LNH System, Nuclear Magnetic Resonance Imaging

IMPRIS Imaging Inc

IMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3 Tesla MRI Magnet, Verio

DateOct 28, 2025 (initiated Sep 30, 2025)
StatusOpen, Classified
ReasonThere is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Root causeUnder Investigation by firm
ActionImris issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 9/30/2025 via email. The notice explained the issue, the risk to health, and provided the following information: "If the error message, "Magnet Supervision: The average heater power is above the alarm threshold. Please contact Siemens Service." appears, DO NOT REBOOT THE SYSTEM. Please call your local Siemens Healthineers service organization. Grant access to the system so that the inspection can be carried out as directed by Siemens Healthineers service organization." "Please immediately instruct your person
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083137NEURO III-SV
Imris, Inc., Dec 16, 2008
K101813IMRX SYSTEM V1 AND IMRISNV/IMRISCARDIO
Imris, Inc., Oct 5, 2010
K133692IMRI 1.5T AND IMRI 3T S
Imris, Inc., Feb 10, 2014
K212367IMRIS iMRI 3T V
Deerfield Imaging, Inc., Aug 23, 2021