Recall Z-0479-2026 · QNI Resorbable Implant For Anterior Cruciate Ligament (Acl) Repair

Miach Orthopaedics

BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;

DateNov 7, 2025 (initiated Feb 2, 2023)
StatusOpen, Classified
ReasonThe field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only.
Root causeUnder Investigation by firm
ActionOn January 30, 2023 RE: BEAR Implant Hydration memo emails were sent to sales staff. If you observe an implant that hydrates too quickly (e.g., less than 30 seconds) and appears that it may not remain intact upon insertion into the knee, do not attempt to implant the device. We would recommend you always have a backup device available to address such circumstances. We plan to revise the Tips and Pearls document, ML-1027, to provide further instructions/details on how to properly hydrate the BEAR device. In the meantime, one way to avoid shortened hydration time is to consider hydrating the de
Quantity167 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K243578BEAR® (Bridge-Enhanced ACL Restoration) Implant
Miach Orthopaedics, Inc., Mar 6, 2025