Recall Z-0155-2026 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

GE Medical Systems, LLC

LOGIQ P10 series with software version R4.5.7 Model Number 5877534

DateOct 15, 2025 (initiated Sep 18, 2025)
StatusOpen, Classified
ReasonThe Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
Root causeSoftware design
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/18/25 was sent to customers. Actions to be taken by Customer/User For LOGIQ P9 and LOGIQ P10 series ultrasound systems with system software version R4.5.7, discontinue using the UGAP feature (please see Appendix for product identification). You may continue using other imaging features. After the correction has been implemented, please destroy the installation media for software R4.5.7 at your site. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please
Quantity65 units
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K214039LOGIQ P10, LOGIQ P9, LOGIQ P8
GE Medical Systems Ultrasound and Primary Care Diagnostics, Mar 22, 2022