Recall Z-0580-2026 ยท MAX Intervertebral Fusion Device With Bone Graft, Lumbar

Medtronic Sofamor Danek USA Inc

Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11

DateNov 21, 2025 (initiated Sep 16, 2025)
StatusOpen, Classified
ReasonExpandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
Root causeLabeling design
ActionMedtronic Sofamor Danek notified consignees on about 09/16/2025 via letter. Consignees were informed of updates to the IFU and surgical guide. Consignees were instructed to share the notification within each facility with clinical users of the affected systems and to complete and return the provided Customer Acknowledgement form.
Quantity275,260 units
DistributionWorldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Netherlands, New Caledonia, New Zealand, Norway, Portugal, Reunion, Saudi Arabia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210425Catalyft PL Expandable Interbody System
Medtronic Sofamor Danek USA, Inc., May 24, 2021