ClearanceBrief
Recall Z-0962-2026 ยท BRY Cabinet, Table And Tray, Anesthesia

CareFusion 303, Inc.

BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD Pyxis CII Safe ES, REF: 1116-00

DateDec 19, 2025 (initiated Nov 20, 2025)
StatusOpen, Classified
ReasonAutomated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.
Root causeDevice Design
ActionOn 11/20/2025, correction notices were mailed and emailed to customers: 1) If a pocket fails with the error message of "Duplicate Address Detected," users with the proper permissions can leverage the "Recover Storage Space" functionality to eject the affected pocket and clear the error. Note: This does not resolve the Firmware v1.48 issue. This only clears the error message from the "Recover Storage Space" screen. Content stored within the affected CUBIE pocket will still need to be manually accessed once the CUBIE pocket is ejected. 2) Facilities should have comprehensive policies and proc
Quantity53669
DistributionWorldwide - US Nationwide distribution including in the states of CO, IN, OH, TX, KS, FL, GA, NC, CA, HI, IL, NY, MI, VA, AL, LA, NE, KY, OK, AR, TN, WY, NJ, WI, SD, AZ, NV, MO, MA, ID, WV, DE, PA, WA, MT, CT, IA, MD, PR, MN, NM, OR, VT, MS, ME, AK, UT, DC, GU, SC, NH, RI, ND, AS and the countries
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.