ClearanceBrief
Recall Z-0376-2026 ยท JAK System, X-Ray, Tomography, Computed

PHILIPS MEDICAL SYSTEMS

Model: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;

DateOct 27, 2025 (initiated Sep 25, 2025)
StatusOpen, Classified
ReasonIssue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software f
Root causeUnder Investigation by firm
ActionOn September 25, 2025, ELECTRONIC PRODUCT RADIATION CORRECTION/URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken. - Follow the instructions within the letter on the use of the systems with updated recommendations. Complete and return the provided response form. Circulate the notice to all users of the device, retain the letter and post it in a visible place likely to be seen/viewed. Actions by recalling firm: Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the solution to resolve the issue
Quantity165 units
DistributionDomestic: AL, AR, CA, CT, FL, GA, IL, IN, KY, MD, MI, OH, PA, PR, TN, TX, UT, VA; International: Argentina, Australia, Austria, Brazil, Czeck Republic, Denmark, Ecuador, Finland, France, Germany, Guyana, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Kenya, Latvia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180015Philips Incisive CT
Philips Healthcare (Suzhou) Co., Ltd., Mar 20, 2018
K212441Philips Incisive CT
Philips Healthcare (Suzhou) Co., Ltd., Apr 27, 2022
K232491CT 5300
Philips Healthcare (Suzhou) Co., Ltd., May 3, 2024