Recall Z-0154-2026 ยท FMG Stopcock, I.V. Set

ICU Medical, Inc.

IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Grad

DateOct 23, 2025 (initiated Oct 2, 2025)
StatusOpen, Classified
ReasonIV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
Root causeProcess control
ActionICU notified consignees via email with an "URGENT MEDICAL DEVICE CORRECTION" on about 10/02/2025. Consignees were instructed to: 1) Check all inventory locations within your institution for the affected products and discontinue use. Destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2) Share this notification with all potential users of the device, to ensure they are aware of this notification. If the devices are used at another location, please
Quantity23203
DistributionWorldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964435PRIMARY IV SET
Icu Medical, Inc., Apr 2, 1997