Recall Z-0490-2026 ยท NLQ Single-Use Reprocessed Ultrasonic Surgical Instruments

Stryker Sustainability Solutions

HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEO, RX ONLY

DateNov 12, 2025 (initiated Sep 24, 2025)
StatusOpen, Classified
ReasonDue to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation.
Root causeProcess control
ActionOn September 23, 2025, Stryker issued an "Urgent: Medical Device Recall" notification dated September 22, 2025, to affected consignees. Styker asked consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product (Attachment B). Segregate the affected products, quarantine and discontinue use. 3. Please complete and return the Business Reply Form to SSSPFA@Stryker.com or use the QR Code on Attachment A. Response is required. a. If you have indicated that product remains
Quantity22, 476 units
DistributionUS Nationwide distribution and OUS (foreign) to country of: Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170456Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology
Stryker Sustainability Solutions, Apr 5, 2017