ClearanceBrief
Recall Z-0416-2026 ยท DQK Computer, Diagnostic, Programmable

Spacelabs Healthcare, Ltd.

Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient an

DateOct 31, 2025 (initiated Sep 16, 2025)
StatusOpen, Classified
ReasonDue to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
Root causeSoftware design
ActionOn 09/15/2025, the firm sent via email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them of two issues associated with the Sentinel Software Version 11.6.0. The issues are: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record. Customers are instructed to: 1. Print and Post Notification: Please print the related Urgent User Notification (Ap
Quantity142
DistributionWorldwide - US Nationwide distribution in the state of MN and the countries of CA, DE, FR, GB, IT, NC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152881SENTINEL CARDIOLOGY INFORMATION MANAGEMENT SYSTEM
Spacelabs Healthcare, Ltd., Apr 1, 2016