Recall Z-0230-2026 ยท HRS Plate, Fixation, Bone

CPM Medical Consultants, LLC.

Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L, OS200135, OS200324, OS200325, OS200326, OS328026, OS328028C, OS216150, OS218150, OS292110, OS292120, OS292160, OS292200, OS200516, OS200518, OS200520, OS200522, OS200616, OS200618, OS200620, OS200622, OS421300, OS421302, OS421304, OS421306, OS421308, OS4

DateOct 21, 2025 (initiated Sep 17, 2025)
StatusOpen, Classified
ReasonInstructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for 5-10 minutes drying time, because these requirements do not apply in the United States.
Root causeLabeling design
ActionOn 9/17/2025, correction notices were mailed to customers and distributors who were informed of the following: - The only sterilization requirement now shown on the new Rev D IFU is 4 mins exposure time at 132-Degrees C (270-Degrees F) and 20-30mins drying time. - All affected versions of the IFU should be immediately discarded. - You can access the e-IFU at: https://fusemedical.com/wp-content/uploads/2025/09/Fuse-ULTRA-Foot-Plating-System-IFU-Rev-D-9.2025.pdf - Please pass on this Medical Device Correction to all users of the products and all other people who need to be aware within your o
Quantity3317
DistributionUS Nationwide distribution in the states of NV, TX, LA, WA, WI, FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141784ORTHO SOLUTIONS ULTOS PLATING SYSTEM
Ortho Solutions Limited, Jul 23, 2014