Recall Z-0420-2026 · PGW Ear, Nose, And Throat Stereotaxic Instrument

Integra LifeSciences Corp. (NeuroSciences)

TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;

DateNov 4, 2025 (initiated Oct 3, 2025)
StatusOpen, Classified
ReasonIt was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.
Root causeDevice Design
ActionOn October 3, 2025 URGENT: VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Actions to be Taken by Customers 1. Examine the your system to determine if you have product on hand subject to this field recall (Rev 00 and Rev01). 2. If you have revision 00 and/or revision 01, please check the box on the enclosed Acknowledgement Form I do have affected product(s) and record the total quantity of the affected product that you have. 3. If you do not have affected product, check the box, I do not have affected product. 4. Complete the rest of the Acknowledgement Form and return. 5.
Quantity1198 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231862TruDi® Navigation System V3 (FG-2000-00)
Acclarent, Inc., Jul 21, 2023