Recall Z-0185-2026 ยท OJK Eye Tray
American Contract Systems Inc.
Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY
| Date | Oct 16, 2025 (initiated Sep 4, 2025) |
| Status | Open, Classified |
| Reason | Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. |
| Root cause | Nonconforming Material/Component |
| Action | Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users. |
| Quantity | 32,433 total |
| Distribution | US Nationwide distribution in the states of SD, IA, MN, WA, IL. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.