Recall Z-0217-2026 ยท JAK System, X-Ray, Tomography, Computed

UIH Technologies LLC

Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only

DateOct 21, 2025 (initiated Sep 12, 2025)
StatusOpen, Classified
ReasonThird-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Root causeOther
ActionOn September 12, 2025, UIH Technologies LLC, issued a recall notification to affected consignees. UIH asked consignees to take the following actions: 1. Ensure patient positioning clearance: When positioning the patient, confirm that arms and legs and clear of all table gaps and moving parts. 2. Avoid use of third-party slickers: We advise against the use of any non-UIH-approved slickers. 3. Service Contact: For technical assistance, contact your United Imaging Service engineer or reach out to our 24/7 Customer Contact Center, Toll-Free: (855) 221-1552/Direct: (832) 699-2799 4.Report ad
Quantity8 units
DistributionUS: OUS:
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181414uCT 530, uCT 550
Shanghai United Imaging Healthcare Co., Ltd., Aug 14, 2018
K200016uCT 530, uCT 550
Shanghai United Imaging Healthcare Co., Ltd., Feb 25, 2020