ClearanceBrief
Recall Z-0173-2026 ยท MLS Labor And Delivery Kit

American Contract Systems Inc.

Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK - 213387 AKDV96B VAGINAL DELIVERY AKDV96B VAGINAL DELIVERY EVBH28D BIRTHING ROOM PACK UPVD01K VAGINAL DELIVERY PACK

DateOct 16, 2025 (initiated Sep 4, 2025)
StatusOpen, Classified
ReasonIdentified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Root causeNonconforming Material/Component
ActionFirm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
Quantity32,433 total
DistributionUS Nationwide distribution in the states of SD, IA, MN, WA, IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.