Recall Z-0188-2026 ยท DQY Catheter, Percutaneous

BALT USA, LLC

Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100, MEGABT091110 Software Version: N/A Product Description: The Mega Ballast 091 is designed with a semi-rigid proximal shaft and a highly flexible distal shaft. The design facilitates the advancement of the Mega Ballast to access the tortuous vasculature over a guid

DateOct 24, 2025 (initiated Sep 25, 2025)
StatusOpen, Classified
ReasonDue to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen.
Root causeProcess control
ActionOn 09/18/2025, the firm sent via Priority Mail an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that, Balt USA has became aware of a situation where multiple Mega Ballast units failed the in-house QC final inspection due to the exterior coating material lifting. Further assessment found this coating also went inside of the ID and protruded from the inner surface of the catheter, which is not intended as part of the manufacturing process. Hospitals are instructed to: - Cease use of the affected product. - Inform hospital staff, such as safety officers, pharmacists, purchas
Quantity207 catheters
DistributionUS distribution to states of: AL, CA, CO, CT, FL, IL, KY, MA, NY, TN, TX, and UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K242376Next Generation Access Platform
Balt USA, LLC, Dec 30, 2024