Recall Z-0216-2026 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Beckman Coulter, Inc.

DxI 9000 Access Immunoassay Analyzer, Part Number C11137

DateOct 21, 2025 (initiated Sep 19, 2025)
StatusOpen, Classified
ReasonBeckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
Root causeSoftware design
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 9/12/25 was sent to customers. " When ordering tests for a manually diluted sample through the Test Order Entry page, a unique sample ID is required. " Alternatively, use the Dilute and Rerun feature or an automated dilution to measure diluted samples. Refer to Rerunning a Test with a Diluted Sample in the System Help for more information. " Beckman Coulter suggests sharing this letter with your Medical Director to determine if performing a retrospective review of results is necessary. " Beckman Coulter is investigating the root cause
Quantity571 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221225Access TSH (3rd IS) Assay, DxI 9000 Access Immunoassay Analyzer
Beckman Coulter, Inc., Nov 10, 2022