ClearanceBrief
Recall Z-0308-2026 ยท NGY Cannula, Trocar, Ophthalmic

Alcon Research LLC

Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000356 27GA HP ENTRY SYSTEM, 4 MM 1 COUNT 8065000358 25GA HP ENTRY SYSTEM, 6 MM 1 COUNT 8065000359 27GA HP ENTRY SYSTEM, 6 MM 1 COUNT 8065000467 25GA INFUSION CANNULA(4&6MM ES)+TRAY+TMD 8065000468 27GA INFUSIO

DateOct 24, 2025 (initiated Aug 5, 2025)
StatusOpen, Classified
ReasonDue to incomplete seals in the pouch which provide the sterile barrier.
Root causeProcess control
ActionOn 08/05/2025, the firm sent via FedEx Overnight an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that for the Alcon Standalone Vitrectomy Consumables, there is potential for some pouches within impacted lots to have an incomplete seal. Customers are instructed to: locate and dispose of any affected lots of Alcon Standalone Vitrectomy Consumables remaining in your inventory. To comply with this Medical Device Recall and request the replacement of any unused product, please take the following steps: 1. Review inventory to determine if there are any unused affected product
Quantity1632
DistributionWorldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,TN, TX, VA, WA, WI, and WV; and OUS (foreign) to countries of: Argentina, Australia, Austria, Azerba
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.