Recall Z-0307-2026 ยท MPA Endoilluminator

Alcon Research LLC

Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000252 UNITY 27 GA CHANDELIER 8065000253 UNITY 27+DS WIDE ANGLE LIGHT PROBE Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures.

DateOct 24, 2025 (initiated Aug 5, 2025)
StatusOpen, Classified
ReasonDue to incomplete seals in the pouch which provide the sterile barrier.
Root causeProcess control
ActionOn 08/05/2025, the firm sent via FedEx Overnight an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that for the Alcon Standalone Vitrectomy Consumables, there is potential for some pouches within impacted lots to have an incomplete seal. Customers are instructed to: locate and dispose of any affected lots of Alcon Standalone Vitrectomy Consumables remaining in your inventory. To comply with this Medical Device Recall and request the replacement of any unused product, please take the following steps: 1. Review inventory to determine if there are any unused affected product
Quantity330
DistributionWorldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,TN, TX, VA, WA, WI, and WV; and OUS (foreign) to countries of: Argentina, Australia, Austria, Azerba
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.