ClearanceBrief
Recall Z-0047-2026 ยท EOQ Bronchoscope (Flexible Or Rigid)

Olympus Corporation of the Americas

OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE 1T60

DateOct 16, 2025 (initiated Sep 11, 2025)
StatusOpen, Classified
ReasonAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Root causeDevice Design
ActionAn URGENT Medical Device Correction notification dated 9/11/25 was mailed to consignees to notify them of this recall action. Consignees are asked to carefully read the recall notification and identify any affected devices in their inventory. Consignees are to ensure all personnel are knowledgeable and aware of the updated IFU. Consignees are to acknowledge receipt of the recall notification through Olympus' web portal at https://olympusamerica.com/recall. Forward the provided recall notification if product was further distributed. Consignees with any questions can contact Cynthia Ow by phone
Quantity347 units US; 2,869 units OUS
DistributionUS Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023984MODIFICATION TO OLYMPUS BRONCHOSCOPES BF-40 SERIES, BF-240 SERIES, AND BF-160 SERIES
The Olympus Optical Co., Jan 31, 2003