Recall Z-2571-2025 ยท RHP Uv Lamp, Germicidal
BigTree Sales Inc.
The product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.
| Date | Sep 18, 2025 (initiated Aug 29, 2025) |
| Status | Open, Classified |
| Reason | The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the accomplishment of the product s primary purpose of disinfecting objects and surfaces, and create a risk of skin and/or eye injuries to user and nearby persons. |
| Root cause | Radiation Control for Health and Safety Act |
| Action | BigTree Sales, Inc. notified FDA and their customers as required by 21 CFR 1003.10(b), that their electronic product, Portable UV-C Light Sanitizer Wand/Travel UV Sanitizing Wand, model: HH1103/HD-UV-01PL03 (SKU BT-GERMWAND-LRG-WHT), has been discovered to contain a defect related to its safety. Their customer letter was sent with the knowledge of the U.S. Food and Drug Administration (FDA). |
| Distribution | U.S.A |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.