Recall Z-0233-2026 ยท FLL Continuous Measurement Thermometer

Baxter Healthcare Corporation

Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers

DateOct 23, 2025 (initiated Sep 17, 2025)
StatusOpen, Classified
ReasonAffected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.
Root causeProcess control
ActionBeginning September 17, 2025, firm notified impacted distributors via Medical Device Recall letters. Customers were instructed to return any affected product and submit a request for replacements. Distributors were instructed to notify their end users via a provided Important Product Information Letter. For general questions regarding this action, please contact Welch Allyn technical service at 800-535-6663. Select the following sequence of options: 2 (technical support), 1(English), 1 (physical exam devices), 3 (thermometry), between the hours of 8:00 AM and 8:00 PM Eastern Time, Monday thro
Quantity5209
DistributionDistribution US nationwide, Canada and Colombia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030580SURETEMP PLUS
Welch Allyn, Inc., May 6, 2003