Recall Z-0125-2026 ยท OGE Epidural Anesthesia Kit
B BRAUN MEDICAL INC
Combined Spinal and Epidural Anesthesia Tray 2, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN 27 Ga. x 5 in. (12.7 cm) Spinal Needle, and 19 Ga. Closed Tip PERIFIX FX Springwound Epidural Catheter. Product Code: SESK. Product Description: SESK COMBINED SPINAL AND EPIDURAL TRAY 2.
| Date | Oct 9, 2025 (initiated Sep 18, 2025) |
| Status | Open, Classified |
| Reason | Potential for the lid of the catheter connector to be in the incorrect position. |
| Root cause | Process control |
| Action | B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distrib |
| Quantity | 60 units |
| Distribution | Nationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.