ClearanceBrief
Recall Z-0166-2026 · DPS Electrocardiograph

Philips North America

Cardiac Workstation 7000; Model Number: 860441;

DateOct 16, 2025 (initiated Sep 12, 2025)
StatusOpen, Classified
ReasonFailure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
Root causeProcess control
ActionOn September 12, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Share this communication with all clinical staff to review and undedrstand. Please this letter with the device. Pass the notice along to all those who need to be aware within the organization or to any organization where affected devices were transferred. Ensure technical solution established by Philips is implemented as soon as available within the timeframe communicated by Philips. In case any device cannot be corrected, inform Philips of the reason why the Technical Solution could no
Quantity29 units
DistributionUS Nationwide distribution in the states of CA, IA, TN, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K241556Cardiac Workstation (5000); Cardiac Workstation (7000)
Philips Medizin Systeme Böblingen GmbH, Dec 17, 2024