Recall Z-0722-2026 ยท IME Pack, Hot Or Cold, Reusable

DJO, LLC

Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02

DateNov 26, 2025 (initiated Sep 25, 2025)
StatusOpen, Classified
ReasonThe IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.
Root causeLabeling Change Control
ActionPer recall notice, customers were asked to do the following: Users should immediately discontinue all further use of this product and discard any opened/used units. If you have unopened, unused units in your possession, please contact firm Customer Service at (844) 331-7767. You will be provided with a return authorization number and shipping label that is required for proper return and identification of the device. All users must complete the acknowledgement and response form and return to productsafety@enovis.com This notice must be distributed to all those within your organization and to
Quantity1,594
DistributionUS Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI, NC, NE, CA, MA, CO, SC, OH, IN, MN, NM, PA, NY, IL, FL, NJ, VA, AZ, VT, CT, OK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall