Recall Z-0122-2026 ยท OGE Epidural Anesthesia Kit

B BRAUN MEDICAL INC

ESPOCAN Combined Spinal and Epidural Anesthesia Tray, ESPOCAN 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN 25 Ga. x 5 in. (12.7 cm) Spinal Needle with Marking and Centering Sleeve, and 19 Ga. Closed Tip PERIFIX FX Springwound Epidural Catheter. Product Code: ES1725KFX. Product Description: ES1725KFX ESPOCAN SPINAL/EPID. TRAY.

DateOct 9, 2025 (initiated Sep 18, 2025)
StatusOpen, Classified
ReasonPotential for the lid of the catheter connector to be in the incorrect position.
Root causeProcess control
ActionB Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distrib
Quantity10,130 units
DistributionNationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
FDA recordOpen on fda.gov