ClearanceBrief
Recall Z-0120-2026 ยท OGE Epidural Anesthesia Kit

B BRAUN MEDICAL INC

PERIFIX FX Continuous Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFS. Product Description: CE17TKFS EPIDURAL TRAY.

DateOct 9, 2025 (initiated Sep 18, 2025)
StatusOpen, Classified
ReasonPotential for the lid of the catheter connector to be in the incorrect position.
Root causeProcess control
ActionB Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distrib
Quantity2,760 units
DistributionNationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.