ClearanceBrief
Recall Z-0208-2026 ยท KNT Tubes, Gastrointestinal (And Accessories)

Cardinal Health 200, LLC

Salem Sump PVC Tubes: Product Code/Product Name: 8888265108 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36 (91 cm) 8888265116 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48 (122 cm) 8888265124 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48 (122 cm) 8888265132 Salem Sump Silicone Dual

DateOct 17, 2025 (initiated Sep 18, 2025)
StatusOpen, Classified
ReasonFirm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Root causeUse error
ActionOn September 18, 2025, the firm notified customers via Urgent Medical Device Product Correction Update letters. This is an update to the notice sent in July 2024 for RES 95214. The present recall includes all lots of additional product codes listed. The letter provides customers with additional guidance on how to properly use the Anti-Reflux Valve (ARV): Important Instructions: After placement of the Salem Sump tube, always inject 10ml-20ml of air before seating the Anti Reflux Valve (ARV) into the blue sump vent lumen. Instructions: 1. Seat the BLUE END of the ARV into the blue sump vent lu
Quantityall lots
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Au
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K830710ARGYLE SILICONE SALEM SUMP TUBE
Sherwood Medical Co., Mar 29, 1983