Recall Z-0071-2026 ยท CEO Phosphomolybdate (Colorimetric), Inorganic Phosphorus

Beckman Coulter Inc.

SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.

DateOct 8, 2025 (initiated Sep 3, 2025)
StatusOpen, Classified
ReasonSpecific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.
Root causeUnknown/Undetermined by firm
ActionOn September 3, 2025, Beckman Coulter issued a "Urgent Medical Device Recall" Notification to affected consignees via E-Mail. Beckman Coulter asked consignees to take the following actions: 1. You may continue to use the affected lots if calibration passes and quality control results are within your facility s acceptance criteria. 2. If the above lots fail calibration due to INIT ADC HI before the lot expiration date, contact Customer Technical Support Center for troubleshooting and/or replacement. 3. If you receive replacement product, review the Safety Data Sheet (SDS) and discard the
Quantity2146 units
DistributionUS: AL KY LA MA MD MS NC NE NJ NY OK PA TX WA WV OUS: Canada Mexico Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.