ClearanceBrief
Recall Z-0562-2026 ยท GIO Tube, Collection, Capillary Blood

ASP Global, LLC. dba Anatomy Supply Partners, LLC.

SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 L; Prepared with Sodium Heparin Model/Catalog Number: 06 0009 Product Description: A clear, plastic capillary blood collection tube

DateNov 19, 2025 (initiated Sep 26, 2025)
StatusOpen, Classified
ReasonDue to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Root causeProcess change control
ActionOn 09/26-29/2025, ASP Global emailed an URGENT MEDICAL DEVICE RECALL Letter to customers. ASP Global, LLC, is voluntarily recalling all RAM Scientific SAFE-T-FILL products within remaining expiry. ASP Global was made aware that our contract manufacturer for the RAM Scientific SAFE-T-FILL products was issued an FDA Warning Letter citing significant violations of the Quality System requirements, including inadequate production and process controls, as well as design controls. Customers are instructed to: a) Discontinue use of the recalled products immediately. b) Quarantine any remaining invent
Quantity160250
DistributionDistribution US nationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.