ClearanceBrief
Recall Z-0549-2026 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

ASP Global, LLC. dba Anatomy Supply Partners, LLC.

AFE-T-FILL Micro Capillary Blood Collection, 200 L; Prepared with Lithium Heparin and Plasma Separator Gel; Green Model/Catalog Number: 07 7220 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a collection device used to draw capillary blood from the finger or heelstick.

DateNov 19, 2025 (initiated Sep 26, 2025)
StatusOpen, Classified
ReasonDue to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Root causeProcess change control
ActionOn 09/26-29/2025, ASP Global emailed an URGENT MEDICAL DEVICE RECALL Letter to customers. ASP Global, LLC, is voluntarily recalling all RAM Scientific SAFE-T-FILL products within remaining expiry. ASP Global was made aware that our contract manufacturer for the RAM Scientific SAFE-T-FILL products was issued an FDA Warning Letter citing significant violations of the Quality System requirements, including inadequate production and process controls, as well as design controls. Customers are instructed to: a) Discontinue use of the recalled products immediately. b) Quarantine any remaining invent
Quantity49000
DistributionDistribution US nationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983517SAFE-T-FILL CAPILLARY BLOOD COLLECTION SYSTEM
Ram Scientific, Nov 12, 1998