Recall Z-0158-2026 ยท BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Fisher & Paykel Healthcare, Ltd.

Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723

DateOct 15, 2025 (initiated Sep 18, 2025)
StatusOpen, Classified
ReasonWhen unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.
Root causeNonconforming Material/Component
ActionOn 9/18/2025, correction notices were communicated via email and phone call. Customers were asked to do the following: 1) Prior to use on a new patient, follow the instructions of the updated Disinfection Kit User Manual. 2) If you have distributed affected devices, notify your customers and include a copy of this notice. 3) Please report all device-related incidents to the firm, distributor or local representative. 3) Complete and return the response form via email to FieldActionUSCA@fphcare.com The Disinfection Kit User Manual updated to include: 7a: Perform Power-Out Alarm test Ensure
Quantity51,144
DistributionUS Nationwide distribution including in the states of TX, MS, LA, NC, TN, NY, FL, IL, UT, MT, OH, MI, NJ, AR, GA, WI, CA, IN, OK, CO, AL, SC, AZ, MD, WA, MO, PA, MA, NH, KY, KS, VA, CT, MN, ND, NM, NE, IA, WV, ID, NV, OR, HI, VT, RI, AL, SD, WY, PR, ME.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131895AIRVO 2 HUMIDIFIER, MYAIRVO 2 HUMIDIFIER
Fisher & Paykel Healthcare, Ltd., Oct 24, 2013