Recall Z-0536-2026 ยท PHX Shoulder Prosthesis, Reverse Configuration

Exactech, Inc.

Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm Model/Catalog Number: 321-59-46 Software Version: N/A Product Description: rTSA Reamer, 46mm Component: Not a component

DateNov 17, 2025 (initiated Sep 5, 2025)
StatusOpen, Classified
ReasonDue to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Root causeRelease of Material/Component prior to receiving test results
ActionOn 09/05/2025, the firm emailed an "URGENT MEDICAL DEVICE RECALL NOTICE" to customers informing that, during routine inventory review, it was discovered that the a number of parts/products, which were pending design verification testing were incorrectly placed in active internal inventory rather than designated quarantine locations. Several part/products were distributed to customers. Customers are instructed to: "Carefully review this notification to ensure that they fully understand the issue and all actions required. "Immediately identify, quarantine, and return any of the subject devices
Quantity1
DistributionU.S.: SC O.U.S.: N/A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042021EXACTECH EQUINOXE SHOULDER SYSTEM
Exactech, Inc., Oct 26, 2004