Recall Z-0108-2026 ยท OUT Intracranial Aneurysm Flow Diverter
MICROVENTION INC.
Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA,
| Date | Oct 10, 2025 (initiated Sep 5, 2025) |
| Status | Open, Classified |
| Reason | Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences. |
| Root cause | Device Design |
| Action | On 9/5/2025, recall notices were mailed to customers who were asked to do the following: 1) Stop using and quarantine affected devices. 2) Provide the notice to medical facilities or users to whom you have distributed recalled devices. 3) Return effected devices. 4) Complete and return the customer acknowledgement form via email to Terumo8943@sedgwick.com Direct questions to the firm at terrence.callahan@microvention.com On 10/29/2025, additional recall notices were distributed identifying additional distributed devices. |
| Quantity | 467 |
| Distribution | Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.