Recall Z-0231-2026 ยท DPS Electrocardiograph

GE Medical Systems, LLC

MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4

DateOct 22, 2025 (initiated Aug 22, 2025)
StatusOpen, Classified
ReasonIf a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.
Root causeDevice Design
ActionAn "URGENT MEDICAL DEVICE CORRECTION" notification letter dated 8/22/25 was sent to customers. Actions to be taken by Customer/User: You can continue to use your device in accordance with the instructions in the MAC VU360 Operator Manual and the actions described below: 1. Inspect all potentially affected devices for the presence of damage to the Acquisition Trunk Cable per the MAC VU360 Operator Manual (P/N 2088531-202, Section 11). 2. If the Acquisition Trunk Cable shows damage, remove the cable from service. 3. Ensure the Acquisition Module is placed correctly in the Acquisition Module Hol
Quantity29,225 units
DistributionWorldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries of: Australia, Austria, Bahrain, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Philipp
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173830MAC VU360 Resting ECG Analysis System
Ge Medical Systems Information Technologies, Inc., Sep 18, 2018