Recall Z-0378-2026 ยท FRN Pump, Infusion

Fresenius Kabi USA, LLC

Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.

DateOct 28, 2025 (initiated Sep 3, 2025)
StatusOpen, Classified
ReasonPotential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation.
Root causeProcess change control
ActionFresenius Kabi notified the consignee on 09/03/2025 via email. The consignee was instructed to review the list of affected units and identify any on hand for return, remove all affected units from circulation pending repair and isolate them to avoid use, notify customers/personnel if the units were further distributed or transferred, and ensure all relevant personnel are aware. A Fresenius Kabi service representative will contact the consignee arrange for the return of affected units for repair.
Quantity6 units
DistributionUS Nationwide distribution in the state of Minnesota.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019