Recall Z-0029-2026 ยท MRN Apparatus, Nitric Oxide Delivery

NOXBOX LTD

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

DateOct 6, 2025 (initiated Sep 9, 2025)
StatusOpen, Classified
ReasonThe unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
Root causeSoftware design
ActionOn September 9, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers: Actions to be taken: - Customers should ensure that all affected personnel are provided with a copy of the recommendations in this letter. - Users should avoid rapidly pressing buttons in quick succession (e.g., go to next screen, accept a value, change a value, backing out, repeating this) as it is necessary to allow an opportunity for the device to respond to user inputs before the next command to avoid a potential software interruption and subsequent restart. - NOxBOX Ltd. is asking customers in the U.S. t
Quantity1667 units
DistributionDomestic: TN;
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171696NOxBOXi Nitric Oxide Delivery System
Praxair Distribution, Inc., Oct 2, 2018
K201339The NOxBOXi Nitric Oxide Delivery System
Praxair, Inc., Jun 19, 2020
K220898NOxBOXi Nitric Oxide Delivery System
Linde Gas & Equipment, Inc., Apr 27, 2022
K231823NOxBOXi Nitric Oxide Delivery System
Linde Gas & Equipment, Inc., Aug 11, 2023
K233251NOxBOXi Nitric Oxide Delivery System
Linde Gas & Equipment, Inc., May 23, 2024