Recall Z-0013-2026 ยท OWB Interventional Fluoroscopic X-Ray System
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Allura Xper FD20/20 OR Table; Model Numbers: 722039; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
| Date | Oct 1, 2025 (initiated Sep 3, 2025) |
| Status | Open, Classified |
| Reason | The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted. |
| Root cause | Device Design |
| Action | On September 3, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. 2. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your syste |
| Quantity | 5 units (OUS only) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmar |
| FDA record | Open on fda.gov |