ClearanceBrief
Recall Z-2634-2025 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Allura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

DateSep 25, 2025 (initiated Sep 3, 2025)
StatusOpen, Classified
ReasonSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Root causeDevice Design
ActionOn September 3, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. "Circulate this Urgent Medical Device Correction to all users of the system so that they are aware of the issue and follow the instructions included in the letter. 2. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this Urgent Medical Device Correction with the documentation of the system u
Quantity1 units (OUS only)
DistributionWorldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Fi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102005ALLURA XPER OR TABLE SERIES
Philips Medical Systems North America, Inc., Aug 9, 2010
K130638ALLURACLARITY XPER FD SERIES X-RAY SYSTEM
Philips Medical Systems Nederland B.V., Jun 28, 2013
K130842ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES
Philips Medical Systems Nederland B.V., May 21, 2013
K133292ALLURA XPER FD; OR TABLE
Philips Medical Systems Nederland B.V., Mar 5, 2014
K141979ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
Philips Medical Systems Netherland BV, Aug 19, 2014
K161563Allura Xper FD series and Allura Xper OR Table series
Philips Medical Systems Nederland B.V., Jul 29, 2016