Recall Z-2629-2025 ยท OWB Interventional Fluoroscopic X-Ray System
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Allura Xper FD20 Biplane OR Table Product Codes: (1)722020, (2) 722025; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
| Date | Sep 25, 2025 (initiated Sep 3, 2025) |
| Status | Open, Classified |
| Reason | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available. |
| Root cause | Device Design |
| Action | On September 3, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. "Circulate this Urgent Medical Device Correction to all users of the system so that they are aware of the issue and follow the instructions included in the letter. 2. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this Urgent Medical Device Correction with the documentation of the system u |
| Quantity | 5 units (2 US, 3 OUS) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Fi |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K102005 | ALLURA XPER OR TABLE SERIES Philips Medical Systems North America, Inc., Aug 9, 2010 |
| K130638 | ALLURACLARITY XPER FD SERIES X-RAY SYSTEM Philips Medical Systems Nederland B.V., Jun 28, 2013 |
| K130842 | ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES Philips Medical Systems Nederland B.V., May 21, 2013 |
| K133292 | ALLURA XPER FD; OR TABLE Philips Medical Systems Nederland B.V., Mar 5, 2014 |
| K141979 | ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES Philips Medical Systems Netherland BV, Aug 19, 2014 |
| K161563 | Allura Xper FD series and Allura Xper OR Table series Philips Medical Systems Nederland B.V., Jul 29, 2016 |