Recall Z-0571-2026 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Greiner Bio-One North America, Inc.

Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product Description: Greiner VACUETTE Blood Collection Tube with Clot Activator and Gel Separator and Standard Stopper (16 x 100 mm, 8ml)

DateNov 19, 2025 (initiated Aug 19, 2025)
StatusOpen, Classified
ReasonSome of the Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat testing. This may result in delayed diagnosis, cause a delay in clinical decision-making or treatment and in some rare cases, yield incorrect results from less stable parameters.
Root causeLabeling Change Control
ActionOn 08/22/2025, the firm sent via email an "Urgent Medical Device Correction" letter to customer informing them that some Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator which may affect the stability of the sample. Customers are instructed to: " Stop use immediately: Discontinue use of the affected product (concerned Item/Lot) and isolate it within your facility. " Notify staff and customers: Inform all relevant employees and notify all customers to whom you distributed the affected products. " Assess risk: Evaluate potential risks or consequences in
Quantity1,980,000
DistributionU.S.: AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, VA and WY. O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081929VACUETTE BLOOD COLLECTION TUBE WITH CLOT ACTIVATOR AND GEL SEPARATOR
Greiner Bio-One North America, Inc., Dec 16, 2008
K983952GREINER VACUETTE BLOOD COLLECTION TUBE
Greiner Meditech, Inc., Dec 28, 1998