ClearanceBrief
Recall Z-2611-2025 ยท NLH Catheter, Recording, Electrode, Reprocessed

Medline Industries, LP

Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbers BD710DF282CRH BD710DF282RRH BD710FJ282CRH BD710FJ282RRH

DateSep 26, 2025 (initiated Jul 31, 2025)
StatusOpen, Classified
ReasonSpecific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Root causeUnder Investigation by firm
ActionOn July 31, 2025, the firm notified customers via first class mail and email of the potential product issue. An updated letter was sent to customers on 8/13/25 with additional item and lot numbers. Affected products are to be returned for credit. Customers were instructed to immediately identify and quarantine all affected product within their possession. Once they have completed and returned the response form listing their affected inventory, they will receive return labels and RGA information. Customers will receive credit when the returned product is received.
Quantity7913 eaches
DistributionDomestic US distribution nationwide. International distribution pending. No international distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K241224Medline ReNewal Reprocessed Biosense Webster Webster CS Catheter with EZ Steer Technology and Medline ReNewal Reprocesse
Surgical Instrument Service and Savings Inc.(Dba Medline Ren, Nov 15, 2024