Recall Z-2600-2025 ยท JAK System, X-Ray, Tomography, Computed

Philips North America Llc

Philips CT systems labeled as the following: 1. Ingenuity CT 728326, Model Number: 728321. 2. Ingenuity CT 728326, Model Number: 728323. 3. Ingenuity CT, Model Number: 728326.

DateSep 19, 2025 (initiated Aug 13, 2025)
StatusOpen, Classified
ReasonThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Root causeProcess control
ActionPhilips Medical notified consignees on about 08/11/2025 via letter. Consignees were instructed that they could continue using the affected systems accordance with the intended use and by following the recommendation: Observe the warning listed in Instructions for Use (IFU) and make sure that the patient is strapped securely to avoid dangling of the hands. Ensure that the patient is placed securely on the patient table and is not in danger of falling. Additionally, consignees were instructed to complete and return the provided response form, to circulate Urgent Medical Device Correction Letter
Quantity285 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauriti
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160743Philips Ingenuity CT