ClearanceBrief
Recall Z-2550-2025 ยท GEH Unit, Cryosurgical, Accessories

Boston Scientific Corporation

ICEfx Cryoablation System, portable helium-free desktop cryoablation unit, Material Numbers (UPN) FPRCH8000, FPRCH8000-02, H749396118000

DateSep 11, 2025 (initiated Aug 18, 2025)
StatusOpen, Classified
ReasonCertain desiccant tube subassemblies were built with end caps that were not correctly tightened.
Root causeProcess control
ActionAn Urgent Medical Device Correction dated 8/18/25 was sent to customers. Customers should cease use of the systems listed in Table 1 and segregate them until a Boston Scientific representative has corrected this issue. A Boston Scientific Field Service Engineer will contact your facility to schedule the correction of your system. For patients in whom an impacted device has already been used, there are no specific recommendations beyond the local standard post-procedure care. If you require additional assistance or more information regarding this communication, please contact your local Bost
Quantity18 units
DistributionWorldwide - US Nationwide and the countries of Canada, France, Germany and Italy. (Preventive Maintenance Kit distributed to Canada only.)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K234002ICEfx Cryoablation System (FPRCH8000-02)
Boston Scientific, Mar 13, 2024