Recall Z-2575-2025 ยท GKZ Counter, Differential Cell
Beckman Coulter Inc.
UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858
| Date | Sep 17, 2025 (initiated Aug 13, 2025) |
| Status | Open, Classified |
| Reason | Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts. |
| Root cause | Component change control |
| Action | On August 13, 2025, Beckman Coulter initiated a "Urgent Medical Device Recall" notification via E-Mail. Beckman Coulter expanded the recall to additional consignees September 25, 2025 Beckman Coulter asked consignees to take the following actions: For specimens with a WBC count > 30 x 103 and = 90 x 103 cells/L, follow the actions listed within the customer letter. For specimens with a WBC count > 90 x 103 cells/L, follow the actions listed within the customer letter. To identify WBC counts > 30 x 103 cells/L, consider implementing decision rules within your Laboratory Information System |
| Distribution | Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Br |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K120771 | UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM Beckman Coulter, Inc., Mar 22, 2013 |