ClearanceBrief
Recall Z-2573-2025 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520

DateSep 17, 2025 (initiated Aug 13, 2025)
StatusOpen, Classified
ReasonHematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
Root causeComponent change control
ActionOn August 13, 2025, Beckman Coulter initiated a "Urgent Medical Device Recall" notification via E-Mail. Beckman Coulter expanded the recall to additional consignees September 25, 2025 Beckman Coulter asked consignees to take the following actions: For specimens with a WBC count > 30 x 103 and = 90 x 103 cells/L, follow the actions listed within the customer letter. For specimens with a WBC count > 90 x 103 cells/L, follow the actions listed within the customer letter. To identify WBC counts > 30 x 103 cells/L, consider implementing decision rules within your Laboratory Information System
DistributionWorldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Br
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K240252UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System; UniC
Beckman Coulter, Inc., Jul 5, 2024