Recall Z-2622-2025 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Mindray DS USA, Inc. dba Mindray North America

BeneVision N1 Patient Monitor (N1). Part Numbers: 6660E-PA00017, 6660E-PA00014, 6660E-PA00012, 6660E-PA00011, 6660E-PA00018, 6660E-PA00015, 6660E-PA00020, Q6660E-PA00011, 6660E-PA00021, 6660E-PA00016, 6660E-PA00019, Q6660E-PA00018, 6660E-PA00010, Q6660E-PA00010, Q6660E-PA00021, Q6660E-PA00012, 6660E-PA00030, 6660E-PA00041, 6660E-PA00035, 6660E-PA00032, 6660E-PA00031. Software Versions: 1.04.00.01,

DateSep 25, 2025 (initiated Aug 15, 2025)
StatusOpen, Classified
ReasonPotential for activation of an abnormal alarm pause.
Root causeSoftware design
ActionMindray contacted consignees via letter on about 08/15/2025. consignees were instructed that they may continue to use the BeneVision N1 patient monitor as a multi-parameter module with Mindray host monitors. Per the additional guidance, it is recommended to confirm that the remaining alarm time counter is actively displayed and follows the existing N1 Operator's Manual: Instructions for when the Alarm Pause is Activated. An image was also included in the notification letter showing the correctly displayed Alarm Pause notification and timer. Consignees were also requested to make all applicab
Quantity2,278 units
DistributionWorldwide - Nationwide distribution and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213799N Series Patient Monitors
Shenzhen Mindray Bio-Medical Electronics Co., Ltd., Jul 12, 2022