Recall Z-2549-2025 · FSM Tray, Surgical, Instrument
Encore Medical, LP
Brand Name: DJO SURGICAL Product Name: FA S Altivate Reverse Glenoid Tray Model/Catalog Number: None Software Version: NA Product Description: Enovis" shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per the Indication for use. While total shoulder replacements are not intended to withstand activity levels and loads of normal healthy bone, they a
| Date | Sep 10, 2025 (initiated Jul 18, 2025) |
| Status | Open, Classified |
| Reason | Their is a potential that the reamer may kick or bind up during or immediately prior to use. |
| Root cause | Device Design |
| Action | On July 29, 2025, Enovis issued a "Urgent Medical Device Recall Notification via Email to affected consignees. On August 4, 2025, Enovis issued an update to the recall notification via E-Mail requesting consignees rework affected trays and returned affected products. Enovis asked consignees to take the following actions: 1. Continue using the wedge reamer assembly while carefully inspecting and following the specific details in the supplied Quality Bulletin. 2. This notice needs to be passed on to all those within your organization or to any organization where the product has been transfer |
| Quantity | 439 kits |
| Distribution | US Nationwide distribution in the states of AL, AZ, CA, CO, FL, HI, IA, IL, IN, KS, KY, LA, MD, ME, NE, NJ, NY, OH, PR, RI, SC, TN, TX, UT, VA, WA, WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K233481 | AltiVate Reverse® Glenoid Encore Medical L.P., May 29, 2024 |