Recall Z-2639-2025 · IYE Accelerator, Linear, Medical

Zap Surgical Systems

ZAP-X Radiosurgery System, Model: 300150

DateSep 26, 2025 (initiated Jul 25, 2025)
StatusOpen, Classified
ReasonIf the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
Root causeSoftware change control
ActionOn 7/25/2025, correction notices were sent to customers who were asked to do the following: Firm will be revising the software to address this issue. Until we supply you with the revised software, the firm strongly advises that you follow these steps to prevent potential collisions. 1. In the rare event that the system triggers a proximity error message during a long gantry move (greater-than-180-degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, please manually extract the patient. 2. After the patient has been safely extracted, please return t
Quantity17
DistributionWorldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211663Zap-X Radiosurgical System
Zap Surgical Systems, Inc., Jul 29, 2021